The Langley ratio method, a new approach for transferring photometer calibration from direct sun measurements
Registro completo de metadatos
Campo DC | Valor | Lengua/Idioma |
---|---|---|
dc.contributor.author | Almansa Rodríguez, Antonio Fernando | es_ES |
dc.contributor.author | Barreto Velasco, África | es_ES |
dc.contributor.author | Kouremeti, Natalia | es_ES |
dc.contributor.author | González, Ramiro | es_ES |
dc.contributor.author | Masoom, Akriti | es_ES |
dc.contributor.author | Toledano, Carlos | es_ES |
dc.contributor.author | Gröbner, Julian | es_ES |
dc.contributor.author | García Cabrera, Rosa Delia | es_ES |
dc.contributor.author | González Ramos, Yenny | es_ES |
dc.contributor.author | Kazadzis, Stelios | es_ES |
dc.contributor.author | Victori, Stéphane | es_ES |
dc.contributor.author | Álvarez, Óscar | es_ES |
dc.contributor.author | Maupin, Fabrice | es_ES |
dc.contributor.author | Carreño Corbella, Virgilio | es_ES |
dc.contributor.author | Cachorro, Victoria E. | es_ES |
dc.contributor.author | Cuevas Agulló, Emilio | es_ES |
dc.date.accessioned | 2024-03-06T08:37:32Z | - |
dc.date.available | 2024-03-06T08:37:32Z | - |
dc.date.issued | 2024 | - |
dc.identifier.citation | Atmospheric Measurement Techniques. 2024, 17(2), 659–675 | es_ES |
dc.identifier.issn | 1867-1381 | - |
dc.identifier.issn | 1867-8548 | - |
dc.identifier.uri | http://hdl.handle.net/20.500.11765/15689 | - |
dc.description.abstract | This article presents a new method for transferring calibration from a reference sun photometer, referred to as the “master”, to a secondary sun photometer, referred to as the “field”, using a synergetic approach when master and field instruments have different spectral bands. The method was first applied between a precision filter radiometer (denoted PFR) instrument from the World Optical Depth Research and Calibration Center (WORCC), considered the reference by the WMO (World Meteorological Organization), and a CE318-TS photometer (denoted Cimel), the standard photometer used by AERONET (AErosol RObotic NETwork). These two photometers have different optics, sun-tracking systems, and spectral bands. The Langley ratio (LR) method proposed in this study was used to transfer calibration to the closest spectral bands for 1 min synchronous data for air masses between 2 and 5, and it was compared to the state-of-the-art Langley calibration technique. The study was conducted at two different locations, Izaña Observatory (IZO) and Valladolid, where measurements were collected almost simultaneously over a 6-month period under different aerosol regimes. | es_ES |
dc.description.sponsorship | Additional funding comes from the European Commission Research Infrastructure Action under the ACTRIS grant (grant no. 871115) and the Ministerio de Economía y Competitividad of Spain through the project SYNERA (grant no. PID2020-521-118793GA-I00). | es_ES |
dc.language.iso | eng | es_ES |
dc.publisher | European Geosciences Union | es_ES |
dc.publisher | Copernicus Publications | es_ES |
dc.rights | Licencia CC: Reconocimiento CC BY | es_ES |
dc.subject | Solar photometry | es_ES |
dc.subject | Aerosol properties | es_ES |
dc.subject | Radiometer | es_ES |
dc.title | The Langley ratio method, a new approach for transferring photometer calibration from direct sun measurements | es_ES |
dc.type | info:eu-repo/semantics/article | es_ES |
dc.relation.publisherversion | https://doi.org/10.5194/amt-17-659-2024 | es_ES |
dc.rights.accessRights | info:eu-repo/semantics/openAccess | es_ES |
Colecciones: | Artículos científicos 2023-2026 |
Ficheros en este ítem:
Fichero | Descripción | Tamaño | Formato | ||
---|---|---|---|---|---|
AMT_Almansa_2023.pdf | 2,71 MB | Adobe PDF | Visualizar/Abrir | ||
amt-17-659-2024-suppl... | 1,09 MB | Adobe PDF | Visualizar/Abrir |