Por favor, use este identificador para citar o enlazar este ítem:
http://hdl.handle.net/20.500.11765/406
Intercomparison of erythemal broadband calibrations performed by AEMET and INTA laboratories
Registro completo de metadatos
Campo DC | Valor | Lengua/Idioma |
---|---|---|
dc.contributor.author | Díaz Rodríguez, Ana María | - |
dc.contributor.author | Postigo González, María | - |
dc.contributor.author | López, María | - |
dc.contributor.author | Vilaplana Guerrero, José Manuel | - |
dc.contributor.author | Serrano Pérez, Antonio | - |
dc.contributor.author | Cancillo, María Luisa | - |
dc.contributor.author | Gröbner, Julian | - |
dc.contributor.author | Hülsen, Gregor | - |
dc.date.accessioned | 2015-12-09T07:50:07Z | - |
dc.date.available | 2015-12-09T07:50:07Z | - |
dc.date.issued | 2012 | - |
dc.identifier.other | NIPO 281-12-009-6 | - |
dc.identifier.uri | http://hdl.handle.net/20.500.11765/406 | - |
dc.description.abstract | National Radiometric Laboratory of AEMET calibrates the AEMET's network of broadband UV radiometers following recommendations of the WMO for this type of calibrations, with a procedure that includes characterisation of each radiometer in the laboratory, utilisation of a radiative transfer model and absolute calibration through a spectral reference instrument using the sun as source. Intercomparisons between several radiometric laboratories are organized periodically for ensuring consistency of whole process of calibration. This technical note describes the first intercomparison between the calibrations at INTA and AEMET laboratories for broadband UV radiometers organised in September 2009. | es_ES |
dc.format | application/pdf | - |
dc.language.iso | eng | es_ES |
dc.publisher | Agencia Estatal de Meteorología | es_ES |
dc.relation.ispartofseries | Nota técnica de AEMET;7 | - |
dc.rights | Licencia CC: Reconocimiento–NoComercial–SinObraDerivada CC BY-NC-ND | - |
dc.subject | Calibración de instrumentos | es_ES |
dc.subject | Radiómetros | es_ES |
dc.subject | Irradiación | es_ES |
dc.title | Intercomparison of erythemal broadband calibrations performed by AEMET and INTA laboratories | es_ES |
dc.type | info:eu-repo/semantics/report | es_ES |
dc.relation.publisherversion | https://doi.org/10.31978/281-12-009-6 | - |
dc.rights.accessRights | info:eu-repo/semantics/openAccess | - |
Colecciones: | AEMET - Publicaciones en línea |
Ficheros en este ítem:
Fichero | Descripción | Tamaño | Formato | ||
---|---|---|---|---|---|
NT AEMET 7 inglés_Int... | 2,16 MB | Adobe PDF | Visualizar/Abrir |
Los ítems de Arcimis están protegidos por una Licencia Creative Commons, salvo que se indique lo contrario.