Please use this identifier to cite or link to this item:
http://hdl.handle.net/20.500.11765/8919
Optical characterisation of three reference Dobsons in the ATMOZ Project – verification of G. M. B. Dobson's original specifications
Title: | Optical characterisation of three reference Dobsons in the ATMOZ Project – verification of G. M. B. Dobson's original specifications |
Authors: | Köhler, Ulf; Nevas, Saulius; McConville, Glen; Evans, Robert; Smid, Marek; Stanek, Martin; Redondas, Alberto
![]() ![]() ![]() |
Keywords: | Dobson spectrophotometers; Instrumental calibration; Total ozone column; Ozone absortion coefficient |
Issue Date: | 2018 |
Publisher: | European Geosciences Union |
Citation: | Atmospheric Measurement Techniques. 2018, 11, p. 1989–1999 |
Publisher version: | https://dx.doi.org/10.5194/amt-11-1989-2018 |
Abstract: | Laboratory investigations of optical characteristics of three standard Dobsons provide real wavelength settings and slit functions. These parameters were compared with the original values given in the manuals. The differences between real and nominal values are not too large, but their application will in any case improve the quality of the total ozone column measurements in the global network. This improvement was the main objective of the ATMOZ project funding these activities. |
Sponsorship : | This work has been supported by the European Metrology Research Programme (EMRP) within the joint research project EMRP ENV59 ATMOZ “Traceability for atmospheric total column ozone”. The EMRP is jointly funded by the EMRP participating countries within EURAMET and the European Union. |
URI: | http://hdl.handle.net/20.500.11765/8919 |
ISSN: | 1867-1381 1867-8548 |
Appears in Collections: | Artículos científicos 2015-2018 |
Files in This Item:
File | Description | Size | Format | ||
---|---|---|---|---|---|
![]() | amt-11-1989-2018.pdf | 1,44 MB | Adobe PDF | ![]() View/Open |
Items in Arcimis are protected by Creative Commons License, unless otherwise indicated.